Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Phys Med Biol ; 68(23)2023 Nov 23.
Artigo em Inglês | MEDLINE | ID: mdl-37934049

RESUMO

Objective. This investigation aims to experimentally determine the charge collection efficiency (CCE) of six commercially available parallel-plate ionisation chamber (PPIC) models in ultra-high dose-per-pulse (UHDPP) electron beams.Approach. The CCE of 22 PPICs has been measured in UHDPP electron beams at the National Metrology Institution of Germany (PTB). The CCE was determined for a dose per pulse (DPP) range between 0.1 and 6.4 Gy (pulse duration of 2.5µs). The results obtained with the different PPICs were compared to evaluate the reproducibility, intra- and inter-model variation, and the performance of a CCE empirical model.Main results. The intra-model variation was, on average, 4.0%, which is more than three times the total combined relative standard uncertainty and was found to be greater at higher DPP (up to 20%). The inter-model variation for the PPIC with 2 mm electrode spacing, which was found to be, on average, 10%, was also significant compared to the relative uncertainty and the intra-model variation. The observed CCE variation could not be explained only by the expected deviation of the electrode spacing from the nominal value within the manufacturing tolerance. It should also be noted that a substantial polarity effect, between 0.914(5) and 1.201(3), was observed, and significant intra- and inter-model variation was observed on this effect.Significance. For research and pre-clinical study, the commercially available PPIC with a well-known CCE (directly measured for the specific chamber) and with a small electrode spacing could be used for relative and absolute dosimetry with a lower-limit uncertainty of 1.6% (k= 1) in the best case. However, to use a PPIC as a secondary standard in UHDPP electron beams for clinical purposes would require new model development to reduce the ion recombination, the polarity effect, and the total standard uncertainty on the dose measurement.


Assuntos
Elétrons , Radiometria , Reprodutibilidade dos Testes , Radiometria/métodos , Frequência Cardíaca , Incerteza
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...